45
000000Ć1150Ć839 IOLMaster 11.02.2004
In this way, complete all selected data records. In the
Data Records
field, the number of data records containing IOL and Post Op Ref data
(first figure) and the total number of loaded data records is specified.
The text boxes below show the number of data records in the
specified axial length ranges.
If you entered all IOL and Post Op Ref data, the requirements for the
optimization calculation have been met.
Note:
If a data record is highlighted in yellow, it does not contain any
ACD values yet:
a0 (Haigis formula) will not be optimized with such data records!
Note:
Data record appearing on a white background contain all data
necesssary for optimization.
Start the optimization calculation by a click on the
OPTIMIZE
button.
Depending on the number of data records to be processed, the
computing process may take some seconds.
Next, the optimized lens constants will be displayed in the new"
column.
Note:
The displayed a0 value does not consider the data records
highlighted in yellow!
Fig. 35 Optimized lens constants
Operation