000000Ć1150Ć839 IOLMaster 11.02.2004
Switching off the IOLMaster
Note:
The procedure described below does not apply if there is a proĆ
blem or if the instrument does not respond to your entries!
If this occurs, immediately switch off the instrument and
disconnect the power cable. Label the instrument as being
defective and call Carl Zeiss Meditec Technical Service.
After you have performed all measurements and calculations, exit
the program using the
Ć appropriate button or the
Ć <
E
> key.
Confirm the operation with
<OK>
or <
Enter
>.
The data of the current (last) patient will be saved automatically.
When Windows has shut down (wait for corresponding message!),
you may switch off the instrument with the power switch (
1
, Fig. 2).
Note:
Failure to switch off the device as described (emergency) may
result in data loss and software damage.
To restart the device, observe the instructions given in Section
What to do in case of problems or system errors,, p. 81
.
Example signal curves of axial length
measurements
Valid signal curves
secondary maxima (systemĆspecific
distance abaout 800
m maximal peak
Ć Very good signals (signalĆtoĆnoise ratio > 10)
Ć Several secondary maxima visible (systemĆspecific)
Ć Clear media, correctly fixating patient
Ć Weak ametropia
Key
Button
E
Operation/
Evaluation of ALM results