84
000000Ć1150Ć839 IOLMaster 11.02.2004
On the
OPTIONS
menu, select
Setup, Add Printer
.
Fig. 77 Printer folder
From the printers window select the icon for the connected printer
(e. g. Epson Stylus Color 880). Right click to view the drop down
menu. Select
Set
as default.
Select the printers not connected (except for the "Add Printer" icon)
and press the <
Del
> key. Follow the instructions displayed on the
screen.
Close the Printer folder again, when the uninstall routine is finished.
If the printer troubles should persist, please call our service represenĆ
tative.
In the United States call Customer Service Carl Zeiss Ophthalmic Division
at 877Ć486Ć7473.
Instrument care
Warning:
Prior to cleaning turn off the instrument and disconnect the power
cable!
Paper pads should be changed and chin rest and forehead rest should
be disnfected after every patient.
For disinfection of parts with which the patient comes into contact
(forehead and chin rest), use a disinfectant authorized for this applicaĆ
tion (see list of disinfectants and methods tested and approved by
Robert-Koch-Institut, 14th edition, of 31 May, 2002).
You can clean all paint surfaces of the equipment with a moist, but not
dripping cloth.
Do not use any aggressive or abrasive cleaning agents.
Wipe off sticky residues using a mild detergent and water.
To clean the display and the keyboard, use commercial cleaning cloths
for computers and monitors.
Caution:
To avoid damage, absolutely take care that no moisture gets into the
instrument or the keyboard.
Care and maintenance