START-UP
Attachments and conversions
Axioskop 2 / Axioskop 2
MOT
2-34
B 40-075 e 06/98
2.2.17
Attachment of intermediate tube
for image projection
The intermediate tube for image projection can
only be used with the binocular tube from the
Axioplan 2 line.
•
Use SW 3 ball-headed screwdriver to loosen
clamping screw (2-47/
8
) and remove available
tube of the Axioskop 2 in an upward direction.
•
Insert dovetail of the adapter for the Axioskop 2
(2-47/
6
) in the stand opening (2-47/
9
) on the
underside and slightly tighten clamping screw
(2-47/
8
).
•
Insert fixation screw (2-47/
5
) in countersunk
hole of the adapter (2-47/
6
) and screw it into
the drilled hole (2-47/
7
) on the stand.
•
Tighten clamping screw (2-47/
8
) and fixation
screw (2-47/
5
).
•
Attach intermediate tube for image projection
(2-47/
2
) on the stand and tighten clamping
screw (2-47/
10
) only slightly at first. Centering
is performed using the dovetail.
•
Insert second clamping screw with washer
(2-47/
3
) in the drilled hole provided on the rear
of the intermediate tube at (2-47/
4
), screw it in
the rear threaded hole of the adapter and
tighten it slightly.
•
Align the intermediate tube with the edges of
the upper stand part and alternately tighten
clamping screws (2-47/
3
) and (2-47/
10
).
•
Use the cover of the tube lens storage case to unscrew the tube lens from the Axioplan 2 tube
(2-47/
1
) and store it in the case.
•
Attach Axioplan 2 tube (2-47/
1
) without tube lens on the intermediate tube, align it with the edges of
the intermediate tube and tighten clamping screw (2-47/
11
).
•
Equip the Axioplan 2 tube (2-47/
1
) with eyepieces with field number 23.
☞
The zoom system always points to the right. The specimen is positioned on this side below the
exit opening of the zoom tube.
10
Fig. 2-47
Attachment of intermediate tube
for image projection
Summary of Contents for Axioskop 2
Page 1: ...Axioskop 2 Routine Microscope Operating Manual...
Page 14: ...INTRODUCTION Axioskop 2 Axioskop 2 MOT 0 14 B 40 075 e 06 98...
Page 120: ...OPERATION Axioskop 2 Axioskop 2 MOT 3 54 B 40 075 e 06 98...
Page 122: ...CARE MAINTENANCE TROUBLESHOOTING AND SERVICE Axioskop 2 Axioskop 2 MOT 4 2 B 40 075 e 06 98...
Page 132: ...CARE MAINTENANCE TROUBLESHOOTING AND SERVICE Axioskop 2 Axioskop 2 MOT 4 12 B 40 075 e 06 98...
Page 134: ...ANNEX Axioskop 2 Axioskop 2 MOT A 2 B 40 075 e 06 98...