START-UP
Initial start-up
Axioskop 2 / Axioskop 2
MOT
2-8
B 40-075 e 06/98
2.1.8
Connection to the line
•
Connect microscope socket (2-9/
2
) to the line
socket via a line cable. The microscope can be
connected to line voltage between 100 and
240 VAC, 60 - 60 Hz. The power unit sets itself
automatically
to the available line voltage
within this range.
•
The HBO 103 illuminator (for epi-fluorescence)
is supplied via a separate transformer. The HBO
100 W transformer must be connected to the
line via its line socket (2-10/
1
).
Before switching on the HBO 100 W
transformer on the instrument rear,
check whether the correct instrument
voltage has been set. If required, the
setting must be changed (see page
4-5). Always pull the power plug
before changing the instrument
voltage!
2.1.9
Switch microscope and HBO 100 W
transformer on and off
•
Switch the microscope on and off using the line
switch (2-9/
1
).
•
Switch the HBO 100 W transformer on and off
using the line switch (2-11/
1
).
Fig. 2-9
Axioskop 2 (rear)
Fig. 2-10
HBO 100 W transformer
(rear)
Fig. 2-11
HBO 100 W transformer (front)
Summary of Contents for Axioskop 2
Page 1: ...Axioskop 2 Routine Microscope Operating Manual...
Page 14: ...INTRODUCTION Axioskop 2 Axioskop 2 MOT 0 14 B 40 075 e 06 98...
Page 120: ...OPERATION Axioskop 2 Axioskop 2 MOT 3 54 B 40 075 e 06 98...
Page 122: ...CARE MAINTENANCE TROUBLESHOOTING AND SERVICE Axioskop 2 Axioskop 2 MOT 4 2 B 40 075 e 06 98...
Page 132: ...CARE MAINTENANCE TROUBLESHOOTING AND SERVICE Axioskop 2 Axioskop 2 MOT 4 12 B 40 075 e 06 98...
Page 134: ...ANNEX Axioskop 2 Axioskop 2 MOT A 2 B 40 075 e 06 98...