START-UP
Axioskop 2 / Axioskop 2
MOT
Attachments and conversions
B 40-075 e 06/98
2-19
2.2.6
Changing the HAL 100 halogen
lamp
•
Switch off the microscope as described in
section 2.1.9, remove plug (2-12/
5
) from 12 V /
100
W socket (2-12/
4
) and allow the
instrument to cool down for approx. 15 mins.
•
Loosen screw (2-29/
5
) until the lamp housing
can be removed in an upward direction.
•
Press both spring levers (2-29/
3
) downwards
and pull out the old halogen lamp (2-29/
2
) in
an upward direction.
•
Press both spring levers downwards, insert new
lamp, with protection cap (2-29/
1
) being
attached in lamp base (2-29/
4
), let go the
spring lever and pull the protection cap off.
•
Press spring lever briefly downwards again to
center the lamp.
•
Attach lamp housing again and tighten
clamping screw (2-29/
5
).
2.2.7
Changing the HBO 103 W/2 mercury pressure short-arc lamp
Before opening the lamp housing, make sure that no connection to the electrical line is
available.
The HBO 203 W/2 lamps may only be changed after they have cooled down. Allow the HBO
103 microscope illuminator to cool down for approx. 15 mins to avoid the risk of burns.
The HBO 103 W/2 lamp may only be removed from the packaging and inserted in the HBO
103 illuminator if a protective mask and safety gloves are worn.
All electrical clamping connections must be made carefully. Pronounced heat during operation
may result in loose contacts.
5
1
2
3
4
Fig. 2-29
Changing the halogen lamp
Summary of Contents for Axioskop 2
Page 1: ...Axioskop 2 Routine Microscope Operating Manual...
Page 14: ...INTRODUCTION Axioskop 2 Axioskop 2 MOT 0 14 B 40 075 e 06 98...
Page 120: ...OPERATION Axioskop 2 Axioskop 2 MOT 3 54 B 40 075 e 06 98...
Page 122: ...CARE MAINTENANCE TROUBLESHOOTING AND SERVICE Axioskop 2 Axioskop 2 MOT 4 2 B 40 075 e 06 98...
Page 132: ...CARE MAINTENANCE TROUBLESHOOTING AND SERVICE Axioskop 2 Axioskop 2 MOT 4 12 B 40 075 e 06 98...
Page 134: ...ANNEX Axioskop 2 Axioskop 2 MOT A 2 B 40 075 e 06 98...