![background image](http://html1.mh-extra.com/html/zeiss/axioskop-2/axioskop-2_manual_3373791047.webp)
START-UP
Axioskop 2 / Axioskop 2
MOT
Attachments and conversions
B 40-075 e 06/98
2-15
2.2
Attachments and conversions
☞
The steps required for attachments to and conversions of the microscope are explained in the
following. After this, the respective modules must be reset in a functioning status again.
2.2.1
Changing the condenser
•
Move stage carrier to the upper stop position
via the focusing drive.
•
Use lever (2-23/
8
) to fold out the front lens on
the available condenser or, in the case of the
motorized version, swing it out by pressing the
respective key.
•
Move condenser carrier (2-23/
2
) to the lowest
position via drive for height adjustment
(2-23/
5
).
•
In the case of the motorized condenser
(Axioskop 2
MOT
), loosen screwing ring (2-24/
1
,
on the left of the stand) of connecting cable
and pull off the connecting cable.
•
Unscrew both centering screws (2-23/
3
) until
condenser can be easily taken out of its mount.
•
Lift available condenser from the orientation
groove by slightly tilting it upwards and pull it
out to the front.
•
Use lever (2-23/
8
) to fold out the front lens on the condenser to be inserted or, in the case of the
motorized version, swing it out by pressing the respective key.
•
Insert condenser (2-23/
1
) between condenser carrier (2-23/
2
) and stage carrier (2-23/
4
). Orient the
locking screw on the underside of the condenser in the direction of the groove (2-23/
7
).
•
Press dovetail of condenser against the spring mount (2-23/
6
) of the condenser carrier until the
condenser can be placed on the condenser carrier horizontally.
•
Let condenser go smoothly; the screw will lock in position in the groove (2-23/
7
) at the front.
•
Screw in centering screws until they engage in the dovetail.
Fig. 2-23
Changing the condenser
Summary of Contents for Axioskop 2
Page 1: ...Axioskop 2 Routine Microscope Operating Manual...
Page 14: ...INTRODUCTION Axioskop 2 Axioskop 2 MOT 0 14 B 40 075 e 06 98...
Page 120: ...OPERATION Axioskop 2 Axioskop 2 MOT 3 54 B 40 075 e 06 98...
Page 122: ...CARE MAINTENANCE TROUBLESHOOTING AND SERVICE Axioskop 2 Axioskop 2 MOT 4 2 B 40 075 e 06 98...
Page 132: ...CARE MAINTENANCE TROUBLESHOOTING AND SERVICE Axioskop 2 Axioskop 2 MOT 4 12 B 40 075 e 06 98...
Page 134: ...ANNEX Axioskop 2 Axioskop 2 MOT A 2 B 40 075 e 06 98...