START-UP
Attachments and conversions
Axioskop 2 / Axioskop 2
MOT
2-30
B 40-075 e 06/98
2.2.15
Attachment of active coobservation equipment for 2 users
The active coobservation equipment for 2 users can only be attached to a Axioskop 2 / Axioskop 2
MOT
microscope stand if the incident-light illuminator (e.g. HBO 103) has been removed.
•
Use SW 3 ball-headed screwdriver to loosen clamping screw (2-45/
12
) and remove available tube in
an upward direction.
•
Remove dust cap (2-45/
14
) from the dovetail at (2-45/
15
) of the coobservation equipment (2-45/
16
).
•
Insert dovetail of coobservation equipment in the stand opening (2-45/
13
) and tighten clamping
screw (2-45/
12
) only slightly at first.
•
Align coobservation equipment with the outer edges of the stand and make sure that the support
points on the coobservation equipment (below) and the stand (2-45/
9
) match exactly.
•
Then tighten clamping screw (2-45/
12
) to prevent the coobservation equipment from shifting.
•
Unscrew the tube lenses from the two binocular tubes (using the cover of one of the storage cases)
and put them in the cases.
•
Remove upper dust caps (2-45/
3
and
18
) from the coobservation equipment.
•
The front dovetail (2-45/
17
) of the coobservation equipment allows attachment of either the
binocular ergotube 6-25°/23 (2-45/
1
) or the ergo-phototube 6-25°/23 (100/100). The rear dovetail
(2-45/
4
) allows not only the attachment of the two ergotubes, but also of the binocular tube 30°/23
(2-45/
2
) or the binocular phototube 30°/23. Then align the tubes and tighten clamping screws
(2-45/
8
and
19
).
The binocular ergotube and the ergo-phototube 6-25°/23 provide an upright and unreversed image of
the specimen in the eyepiece.
•
Screw heat protection shield (2-45/
6
) to coobservation equipment from below using two fixation
screws (2-45/
7
).
•
Insert plug (2-45/
11
) of power unit (2-45/
10
, 115 V or 230 V) in the socket (2-45/
5
) on the
coobservation equipment and tighten the screwing ring to prevent shifting.
•
Connect power unit to the line (115 V or 230 V).
Summary of Contents for Axioskop 2
Page 1: ...Axioskop 2 Routine Microscope Operating Manual...
Page 14: ...INTRODUCTION Axioskop 2 Axioskop 2 MOT 0 14 B 40 075 e 06 98...
Page 120: ...OPERATION Axioskop 2 Axioskop 2 MOT 3 54 B 40 075 e 06 98...
Page 122: ...CARE MAINTENANCE TROUBLESHOOTING AND SERVICE Axioskop 2 Axioskop 2 MOT 4 2 B 40 075 e 06 98...
Page 132: ...CARE MAINTENANCE TROUBLESHOOTING AND SERVICE Axioskop 2 Axioskop 2 MOT 4 12 B 40 075 e 06 98...
Page 134: ...ANNEX Axioskop 2 Axioskop 2 MOT A 2 B 40 075 e 06 98...