![background image](http://html1.mh-extra.com/html/zeiss/axioskop-2/axioskop-2_manual_3373791041.webp)
START-UP
Axioskop 2 / Axioskop 2
MOT
Initial start-up
B 40-075 e 06/98
2-9
2.1.10
HAL 100 halogen illuminator
(1)
Attachment of HAL
100 halogen
illuminator
•
Remove cover from transmitted-light tube
(2-12/
3
).
•
Insert dovetail (2-13/
2
) of lamp housing (2-12/
1
)
in transmitted-light tube (2-12/
3
) on the
instrument rear and use the SW 3 ball-headed
screwdriver to tighten clamping screw (2-12/
2
).
•
Connect 3-pin lamp plug (2-12/
5
) to 3-pin
12 V/100 W socket (2-12/
4
) on the instrument
rear.
(2)
Adjustment of HAL
100 halogen
illuminator
a)
Coarse adjustment
•
Loosen clamping screw (2-12/
2
) and remove
the halogen illuminator from the microscope
stand.
•
Switch on microscope as described under 2.1.9.
•
Direct the light beam against a projection area
(wall) approx. 3 m away.
Do not look into the light exit opening
of the illuminator.
•
Set adjusting screw (2-13/
3
) using SW 3 ball-
headed screwdriver in such a way that both
images of the lamp coil are as sharply focused
as possible on the projection screen.
•
Then set adjusting screws (2-13/
4
and
5
) in
such a way that the lamp coil of one image
exactly fills the gaps in the reflector image
(2-13/
1
).
Fig. 2-12
Attachment of HAL 100 halogen
illuminator
Fig. 2-13
Adjustment of HAL 100 halogen
illuminator
Summary of Contents for Axioskop 2
Page 1: ...Axioskop 2 Routine Microscope Operating Manual...
Page 14: ...INTRODUCTION Axioskop 2 Axioskop 2 MOT 0 14 B 40 075 e 06 98...
Page 120: ...OPERATION Axioskop 2 Axioskop 2 MOT 3 54 B 40 075 e 06 98...
Page 122: ...CARE MAINTENANCE TROUBLESHOOTING AND SERVICE Axioskop 2 Axioskop 2 MOT 4 2 B 40 075 e 06 98...
Page 132: ...CARE MAINTENANCE TROUBLESHOOTING AND SERVICE Axioskop 2 Axioskop 2 MOT 4 12 B 40 075 e 06 98...
Page 134: ...ANNEX Axioskop 2 Axioskop 2 MOT A 2 B 40 075 e 06 98...