START-UP
Axioskop 2 / Axioskop 2
MOT
Attachments and conversions
B 40-075 e 06/98
2-27
2.2.11
Attachment of adapter for the
Axioskop 2
The adapter for the Axioskop 2 is used to attach
intermediate tubes and/or tubes of the Axioplan 2
to the Axioskop 2 / Axioskop 2
MOT
.
•
Use SW 3 ball-headed screwdriver to loosen
clamping screw (2-41/
5
) and remove available
tube of the Axioskop 2 in an upward direction.
•
Insert dovetail of the adapter for the Axioskop 2
(2-41/
3
) in the stand opening (2-41/
6
) on the
underside and slightly tighten clamping screw
(2-41/
5
).
•
Insert fixation screw (2-41/
2
) in countersunk
hole of the adapter (2-41/
3
) and screw it into
the drilled hole (2-41/
4
) on the stand.
•
Tighten clamping screw (2-41/
5
) and fixation
screw (2-41/
2
).
•
Attach required intermediate tube and/or tube (2-41/
1
) of the Axioplan 2 to the adapter and clamp it
using screw (2-41/
7
) (also see the Axioplan 2 manual).
•
Insert eyepieces with field number 23 into the Axioplan 2 tube (2-41/
1
).
Fig. 2-41
Attachment of adapter for the
Axioskop 2
Summary of Contents for Axioskop 2
Page 1: ...Axioskop 2 Routine Microscope Operating Manual...
Page 14: ...INTRODUCTION Axioskop 2 Axioskop 2 MOT 0 14 B 40 075 e 06 98...
Page 120: ...OPERATION Axioskop 2 Axioskop 2 MOT 3 54 B 40 075 e 06 98...
Page 122: ...CARE MAINTENANCE TROUBLESHOOTING AND SERVICE Axioskop 2 Axioskop 2 MOT 4 2 B 40 075 e 06 98...
Page 132: ...CARE MAINTENANCE TROUBLESHOOTING AND SERVICE Axioskop 2 Axioskop 2 MOT 4 12 B 40 075 e 06 98...
Page 134: ...ANNEX Axioskop 2 Axioskop 2 MOT A 2 B 40 075 e 06 98...