START-UP
Initial start-up
Axioskop 2 / Axioskop 2
MOT
2-10
B 40-075 e 06/98
b)
Fine adjustment
•
Attach illuminator to the stand again and tighten clamping screw (2-12/
2
).
•
Remove diffusion disk (3-1/
15
) and filters (3-1/
18
), if switched in, from the beam path.
•
Focus on the specimen using objective
≤
40x and search for a free object area.
•
Remove eyepiece and center lamp coil and its reflector image in the pupil image using adjusting
screws (2-13/
4
and
5
).
•
Use adjusting screws (2-13/
3
) to optimize the homogeneous illumination of the pupil image.
•
Switch in diffusion disk and required filters.
2.1.11
HBO 103 illuminator
(1)
Attach HBO 103 W/2 mercury vapor
short-arc lamp
The HBO 103 illuminator and the HBO 103 W/2
mercury vapor short-arc lamp are supplied in
separate packages for safety reasons.
Therefore, insertion of the HBO 103 W/2 lamp into
the lamp housing is the first step to the start-up of
this illuminator.
Insertion of a new HBO 103 W/2 lamp is described
in detail on page 2-19, section 2.2.7 entitled
"Changing the HBO 103 W/2 mercury vapor short-
arc lamp".
(2)
Attachment of HBO 103 illuminator
•
Remove cover from the incident-light tube
(2-14/
1
).
•
Insert dovetail of lamp housing in incident-light
tube (2-14/
1
) on the instrument rear and use
SW
3 ball-headed screwdriver to tighten
clamping screw (2-14/
2
).
Fig. 2-14
Attachment of HBO 103 illuminator
Summary of Contents for Axioskop 2
Page 1: ...Axioskop 2 Routine Microscope Operating Manual...
Page 14: ...INTRODUCTION Axioskop 2 Axioskop 2 MOT 0 14 B 40 075 e 06 98...
Page 120: ...OPERATION Axioskop 2 Axioskop 2 MOT 3 54 B 40 075 e 06 98...
Page 122: ...CARE MAINTENANCE TROUBLESHOOTING AND SERVICE Axioskop 2 Axioskop 2 MOT 4 2 B 40 075 e 06 98...
Page 132: ...CARE MAINTENANCE TROUBLESHOOTING AND SERVICE Axioskop 2 Axioskop 2 MOT 4 12 B 40 075 e 06 98...
Page 134: ...ANNEX Axioskop 2 Axioskop 2 MOT A 2 B 40 075 e 06 98...