START-UP
Attachments and conversions
Axioskop 2 / Axioskop 2
MOT
2-16
B 40-075 e 06/98
•
In the case of the motorized condenser
(Axioskop 2
MOT
), plug connecting cable in the
socket on the left front of the stand (2-24/
1
)
and tighten screwing ring.
•
The achromatic-aplanatic condenser 1.4 H DIC
and other condensers from the Axioplan 2 line
are also suitable for attachment.
2.2.2
Changing the phase stop for phase
contrast or darkfield in the
universal condenser (if required)
•
Tighten both centering screws (2-25/
1
) until
stop using SW 1.5 Allen screwdriver.
•
To loosen the cover on the condenser underside
(2-25/
2
), loosen both grub screws and remove
the cover. Position turret disk with phase stop
to be changed in the exchange opening and
hold it on the knurled ring, since no click-stop is
effective in this position.
•
Use mounting device contained in tool set to
unscrew retainer ring (2-25/
4
) from the
condenser underside (2-25/
2
) and allow phase
stop (2-25/
3
) to slide out.
•
To insert the new phase stop, turn condenser round, insert the new phase stop, tighten retainer ring,
loosen both centering screws and recenter phase stop using the centering telescope. Make sure that
the correct label is visible on the knurled ring of the turret when the unit is swung into the beam
path.
Fig. 2-24
Attachment of condenser
MOT
1
Fig. 2-25
Changing the phase stop
Summary of Contents for Axioskop 2
Page 1: ...Axioskop 2 Routine Microscope Operating Manual...
Page 14: ...INTRODUCTION Axioskop 2 Axioskop 2 MOT 0 14 B 40 075 e 06 98...
Page 120: ...OPERATION Axioskop 2 Axioskop 2 MOT 3 54 B 40 075 e 06 98...
Page 122: ...CARE MAINTENANCE TROUBLESHOOTING AND SERVICE Axioskop 2 Axioskop 2 MOT 4 2 B 40 075 e 06 98...
Page 132: ...CARE MAINTENANCE TROUBLESHOOTING AND SERVICE Axioskop 2 Axioskop 2 MOT 4 12 B 40 075 e 06 98...
Page 134: ...ANNEX Axioskop 2 Axioskop 2 MOT A 2 B 40 075 e 06 98...