START-UP
Axioskop 2 / Axioskop 2
MOT
Initial start-up
B 40-075 e 06/98
2-13
•
If required, set knurled knob (2-17/
6
) and adjusting screw (2-17/
4
) alternately until both focal spots
have the same dimension (see Fig. 2-19).
•
Use adjusting screws (2-17/
1
and
3
) to adjust the brighter focal spot in the adjusting circle, e.g. on
the left side, in the same way as the focal spot image (2-20).
•
Use adjusting screws (2-17/
2
and
5
) marked in red to adjust the dark focal spot in the adjusting circle,
e.g. on the right side, in the same way as the focal spot image (2-17/
7
and 2-19).
•
Push in adjusting aid again.
☞
The two focal spots of the HBO 103 W/2 lamp must
not
superimpose in the adjusting circle of
the adjusting aid!
If required, the above settings for fine adjustment of the HBO 103 illuminator can be
performed repeatedly.
Fine adjustment without adjusting aid
•
Attach HBO 103 illuminator to microscope
stand as described in section 2.1.11 (2).
•
Remove one objective from the nosepiece and
switch empty opening in function position.
•
Place a white sheet of paper on the mechanical
stage.
•
Fully open the precentered luminous-field and
aperture diaphragms in the incident-light
illumination beam path.
•
The pupil image with the two focal spot images
is now visible on the white sheet of paper.
If required, reduce light intensity using
attenuation filters in the 3-position or
6-position filter slider.
•
Use knurled knob (2-17/
6
) to focus on the
brighter focal spot via collector adjustment.
•
Apply SW 3 ball-headed screwdriver on adjusting screw (2-17/
4
) to focus on the slightly darker,
reflected focal spot of the lamp.
•
If required, set knurled knob (2-17/
6
) and adjusting screw (2-17/
4
) repeatedly until both focal spots
have the same dimension.
•
Use adjusting screws (2-17/
1
and
3
) to set the brighter focal spot in accordance with Fig. 2-21.
•
Use adjusting screws (2-17/
2
and
5
) marked in red to set the darker focal spot in accordance with
Fig. 2-21.
Fig. 2-21
Focal spot imaging without
adjusting aid on the mechanical
stage
Summary of Contents for Axioskop 2
Page 1: ...Axioskop 2 Routine Microscope Operating Manual...
Page 14: ...INTRODUCTION Axioskop 2 Axioskop 2 MOT 0 14 B 40 075 e 06 98...
Page 120: ...OPERATION Axioskop 2 Axioskop 2 MOT 3 54 B 40 075 e 06 98...
Page 122: ...CARE MAINTENANCE TROUBLESHOOTING AND SERVICE Axioskop 2 Axioskop 2 MOT 4 2 B 40 075 e 06 98...
Page 132: ...CARE MAINTENANCE TROUBLESHOOTING AND SERVICE Axioskop 2 Axioskop 2 MOT 4 12 B 40 075 e 06 98...
Page 134: ...ANNEX Axioskop 2 Axioskop 2 MOT A 2 B 40 075 e 06 98...