Instruction Manual 

 

 
 

SMZ Series

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
Copyright © 02/02 
Motic Microscopes, European Division. 

Summary of Contents for SMZ-140-N2GG

Page 1: ...Instruction Manual SMZ Series Copyright 02 02 Motic Microscopes European Division ...

Page 2: ... Support collar screw 18 Focus knob 1 Eyepiece 14 Head 2 Diopter adjuster 3 Eyepiece tubes 5 Head holder locking screw 17 Head holder 15 Power cord for incident illumination 4 Prism housing 13 Base 8 Incident light adjustment screw 6 Objectives 7 Incident illumination 9 On Off switch 12 Stage plate locking screw 11 Stage 16 Zoom knob SMZ 140 N2GG 2 ...

Page 3: ...nents of the stereomicroscope have been carefully packed to make sure they reach you in perfect condition We recommend that you do not discard any packing containers in case you need to return the microscope or store it for long periods of time or should it become necessary to transport it to a technical service for any repair or maintenance procedure The box should contain the following component...

Page 4: ...ings etc The black and white stage is used for non transparent objects or for dissection For best contrast choose the side of the stage to use Warning Transmitted illumination can ONLY be used with the frosted glass stage THE HEAT GENERATED BY THE TRANSMITTED ILLUMINATION CAN MELT OR DAMAGE THE BLACK AND WHITE STAGE SUCH DAMAGE WOULD NOT BE INCLUDED UNDER WARRANTEE 1 Changing the stage a Loosen th...

Page 5: ...entre of the stage 11 3 Turn focussing knobs 18 to mid focus range 4 The head holder 17 is mounted on a column 21 on which it can be moved up or down depending on the size of the object to be focussed on a Support head holder 17 with one hand without touching any lens and with the other loosen the screw 20 on the support collar 19 The head holder can then be slid to the base 13 b Without letting g...

Page 6: ...extended position the image is transmitted to the vertical port but the image cannot be seen by the right eyepiece A To adapt a photographic camera an adapter tube is needed This tube includes 2X lenses to balance the correct parfocality between the images received by the binocular and vertical ports This balance can be achieved providing that the parfocality has been adjusted correctly as describ...

Page 7: ...that the CS mount is responsible Place or remove it according to the procedure to obtain parfocality Maintenance WARNING FOR YOUR OWN SAFETY SWITCH OFF AND DISCONNECT THE MICROSCOPE FROM ANY ELECTRICAL SOURCE BEFORE ATTEMPTING ANY MAINTENANCE PROCEDURE TO AVOID THE RISK OF ELECTROCUTION CONSULT YOUR DISTRIBUTOR IF ANY REPAIR OR MAINTENANCE PROCEDURE IS REQUIRED TO YOUR MICROSCOPE THAT DOES NOT APP...

Page 8: ...t illumination bulb a Unscrew the illuminator protector tube 7 turning it anti clockwise and removing the tube from that of the lenses b With a cloth carefully pull out the bulb to disconnect it from the socket g If the bulb is accidentally touched with bare hands it must be cleaned as this could affect the transmission of light and life span of the bulb c Replace illuminator lens tube turning in ...

Page 9: ...ely as possible without the head sliding down with its own weight The tension adjustment collar for focussing Fig 1 is situated between the focussing knob 18 and the head holder 17 a Loosen the screw located in the collar hole with the 2mm hexagonal key b To increase the tension turn the collar anti clockwise to decrease it turn the collar clockwise c Re tighten the hexagonal key 10 ...

Page 10: ...Fuse cover badly closed Bad connection with power source Insert bulb correctly Replace bulb Close correctly Have repaired by a qualified specialised technician Fuse blown quickly Wrong fuse Replace with appropriate fuse Fuse blown instantly Short circuit Have repaired by a qualified specialised technician IMAGE QUALITY PROBLEM CAUSE SOLUTION Poor resolution Eyepieces dirty Objectives dirty Clean e...

Page 11: ...ails of the required revision Warrantee All MOTIC microscopes are warranted against any manufacturing defect for a 5 year period Damage occurring by any unauthorised repair work or occurring through misuse or modification of the microscope will not be included under the conditions of the warrantee Bulbs and fuses are not under warrantee The warrantee service is provided by MOTIC or its authorised ...