9-16
IM AQ6376-01EN
No. Message
Cause of Warning
No.300 - 399: Errors during the execution of program functions
300
Parameter out of range
A variable value is out of range or is not defined for a command that
sets a parameter using variables.
302
Scale unit mismatch
There is a difference between the Y-axis scale of the active trace
and the unit of a parameter in the “LINE MKR 3 or 4” command.
303
No data in Active trace
Setting of the moving marker, a peak (or bottom) search, or
activation of the analysis function was made with no data in the
active trace.
304
Marker value out of range
Specified wavelength was out of the sweep range in the moving
marker or line wavelength marker setting command.
305
No data in trace A or B
No waveform data in traces A or B when executing the “EDFA NF”
command
306
Invalid data
Trace had no data when attempting to save it to memory or to write
it to FD/HDD.
307
Unsuitable Write item
All data items were OFF at execution of “WRITE DATA”.
320
Undefined variable
A command containing an undefined variable was executed.
321
Variable unit mismatch
The unit of each variable does not agree within a command
containing two or more variables.
322
Overflow
An overflow occurred in an arithmetic operation.
323
Undefined marker variable
A command containing a marker-value variable was executed when
no marker had been displayed.
324
Invalid marker variable
A command containing the corresponding variable was executed at
a time other than immediately after execution of a spectrum width
search, peak search, etc.
325 Undefined line number
GOTO command’s jumping destination is a number other than 1 to
200.
326 F1 greater than F2
F1>F2 when the “IF F1 ≤ @@@@@ ≤ F2” command was executed.
340
Printer paper empty
No printer paper
341
Printer head up
No print is made because the printer’s head-up lever is raised.
345
Option does not respond
No response from an external device.
346
Option is not connected
No external device is connected.
347
GPIB2 not system controller
System controller connected to the GP-IB2 port has been set to an
external computer.
360
Disk full
No file can be created due to insufficient free space on the USB
storage medium.
361
USB Storage not inserted
USB storage medium not inserted.
362
USB Storage is write protected
The USB storage device is write protected.
363
USB Storage not initialized
USB storage not initialized. Or, it has been formatted in a format not
supported by this instrument.
364
Directory full
Directory is full, therefore no file can be created.
365
File not found
The specified file cannot be read because it has not been found. Or,
the file does not exist on the disk.
366
File is write protected
The file is specified to be read only, so that it cannot be rewritten or
deleted.
367
No data
No data to store.
368
File is not a trace file
A file cannot be read because it is not a trace file.
369
Illegal file name
A file cannot be saved due to an incorrect file name.
380
Undefined program
An attempt was made to run a program that is not defined.
381
Syntax error
Command incorrect (a program has been rewritten for some
reason).
9.10 Warning Display Function