1-7
IM AQ6376-01EN
Functions
1
2
3
4
5
6
7
8
9
10
11
App
Index
1.3 Waveform Display
Zoom
<<See section 5.1 for the operating procedure>>
This function allows you to zoom freely in and out on measured waveforms.
You can easily zoom an area simply by selecting it with the mouse. Drag the mouse over
an area of the displayed waveform to zoom in on the area. ZOOMING is displayed at the
bottom of the screen to indicate that the screen is displaying a zoomed area. You can
also zoom in and out with key operations.
Overview
<<See section 5.1 for the operating procedure>>
This function displays an overview window at the very bottom of the waveform display
area. The overview window is displayed when the waveform display is enlarged or
reduced using the zoom function. (Only displayed when a zoom is performed.)
You can easily check which part of the measured waveform is zoomed. The overview
window can be shown or hidden, and you can change its position and size.
Trace
<<See section 4.12 for the operating procedure>>
A trace shows a waveform and measurement conditions. The instrument has a total of
seven independent traces (A through G). Multiple traces can be displayed at the same
time on the waveform screen. In addition, display ON/OFF and mode settings can be set
separately for each trace.
Traces can be set to the following modes.
• WRITE mode
Section 5.2
•
FIX mode
Section 5.2
•
MAX/MIN HOLD mode
Section 5.3
•
ROLL AVG mode
Section 5.4
•
CALCULATE mode
Section 5.5
The following explains each of the modes.
WRITE Mode
Waveform data are written during sweeping.
When a trace is set to WRITE mode, waveform data are written during measurement,
and the data are updated. Traces used in measurements are normally set to WRITE
mode. The trace display at the side of the data area changes to “WRITE.”
FIX Mode
Fixes the data; does not write waveform data.
When a trace is set to FIX mode, its waveform data are not overwritten even when
measurement is performed. Therefore, the waveform on the screen is not overwritten. If
you want to fix the waveform data of a trace, set the trace to FIX mode. The trace display
at the side of the data area changes to “FIX.”