9-14
IM AQ6376-01EN
No. Message
Cause of Warning
111
<G=MKR FIT> failed
Trace G cannot write fitting curve because numbers of data is not
sufficient during <G=MKR FIT>.
120
USB Storage not inserted
USB storage medium not inserted.
121
USB Storage not initialized
SUB storage not initialized.
122
USB Storage is write protected
The USB storage device is write protected.
123
File not found
The specified file cannot be read because it has not been found. Or,
the file does not exist on the disk.
124
Illegal directory name
Directory cannot be created because the name is not valid.
125
Illegal file name
A file cannot be saved due to an incorrect file name.
126
Directory already exists
Directory cannot be created because a directory of the same name
already exists.
128
File is write protected
The file cannot be re-written or deleted because its property has
been set to READ ONLY.
129
USB Storage full
The file cannot be saved, because the hard disk or the USB storage
device is full.
130
Directory full
No files can be created because the directory area is full.
131
No data
There was no data to be saved, although an attempt was made to
save the file.
132
File is not a trace file
The file cannot be read because it is not a waveform file.
133
File is not a program file
The file cannot be read because it is not a program file.
134
File is not a data file
The file cannot be read because it is not a data file.
135
File is not a settings file
The file cannot be read because it is not a settings file.
137
File is not a template file
File cannot be loaded since it is not a template file or there is a
problem with the format of the template data.
138
Cannot copy
The file copy cannot be executed because the “copy from” file and
the “copy to” file have the same name.
140
No paste possible
The paste operation cannot be executed during the editing of
programs because the number of blank rows is not sufficient.
141
No merge possible
It is not possible to execute a merge during the editing of programs
because the merged result would exceed the maximum number of
rows.
142
WL calibration failed
Calibration cannot be executed because the level of a light source
is not sufficient during the wavelength calibration or because
wavelength deviations exceed the calibration range.
143 Optical Alignment failed
Alignment adjustment cannot be executed because the level of a
light source is not sufficient during the alignment adjustment.
144 Go/No go judgment stopped
Go/No go judgment stopped because template data reading or
AUTO sweep was carried out with Go/No go judgment function set
to ON.
145 No template data
Go/No go judgment or template data display was attempted to be
made with no template data provided.
150
File is not a logging file
The file cannot be loaded because it is not a logging file.
151
Disk space is not enough for logging
Logging cannot be started because there is not enough space for
saving waveform files with the logging function.
152
Logging was skipped for Auto zeroing
Unable to log at the specified interval because auto offset was in
progress at the time logging would have occurred (displayed only
the first time after logging is started).
153
Sweep time exceeds the set interval
Unable to log at the specified interval because the sweep time is
longer than the interval (displayed only the first time after logging is
started).
154
This file is not compatible
The file could not be loaded because the model does not support it.
160
Printer paper empty
There was no printer paper, although an attempt was made to use
the internal printer for printing.
161
Printer head up
It was not possible to execute printing because the head-up lever of
the internal printer was raised.
164
Printer Head Temperature error
The temperature of the built-in printer’s printer head is abnormal
when printing.
170
Illegal character
An illegal character was entered for a network name.
9.10 Warning Display Function