6-50
IM AQ6376-01EN
Template Data
• Template data consist of wavelength and level data. Up to 50,001 points of data may
be defined.
• An upper limit, lower limit, and target line can be set.
• The on-screen template data display range and the Go/No Go test function execution
range follow the display scale wavelength range.
Relationship between Go/No Go Judgment and Wavelength Range
-70
-50
-30
-10
10
1547.5
1550
1552.5
REF
FAIL
Upper Limit Line
Lower Limit Line
Le
ve
l (
dB
m
)
Wavelength (nm)
NG points
-70
-50
-30
-10
10
1549
1550
1551
REF
PASS
Upper Limit Line
Lower Limit Line
Le
ve
l (
dB
m
)
Wavelength (nm)
If the waveform is zoomed and the
NG point is outside the displayed
range, the judgment result is PASS.
Note
When performing Go/No Go judgment, display the wavelength range on the screen. The Go/No
Go testing process is not performed on wavelength ranges that are not displayed.
6.12 Go/No-Go Judgment (Template)