App
Index
Explanation
HOLD
The Hold function is used when displaying measured waveforms with the upper and
lower portions of the screen having different wavelength ranges.
For example, after measuring trace A in the upper screen, you can hold the upper screen,
change measurement conditions, then measure trace B in the lower screen.
HOLD has the characteristics below. These are the same for the upper and lower splits.
• The display scale is fixed.
• The trace is fixed.
• When HOLD is applied to a screen with the active trace (WRITE @), the active
trace automatically changes to the FIX state. (FIX @)
• When a screen setting is changed from the HOLD state to NORMAL DISPLAY, the
last set display scale is set as the display scale.
• When a trace in the HOLD state (FIX state) is set to a state other than FIX,
the HOLD is cleared automatically. When this happens, a warning message is
displayed.
5.10 Displaying a Split Screen