No. Message
Cause of Warning
171
Illegal address
An illegal address was set to an IP address.
172
This version is not compatible
Attempted to load an incompatible update file.
173
Update file read error
An incorrect number of files or file corruption was found when
attempting to load the update file.
174
Invalid password
The password entered for the key lock is not correct.
175
Re-entered password is not matched
The two passwords entered for the key lock do not match.
176
Password needs to be 4-digit number
The password entered for the key lock is not a 4-digit number.
No. 200 - 299: Warnings for hardware failures
200
Fan motor stopped!
The fan motor (main unit) stopped. Automatically shuts down after
10 seconds from occurrence.
201
Calibration data failed!
Started up in emulation mode because there was a problem with the
instrument’s calibration data.
202
Fan motor stopped
The CPU fan stopped.
205
Internal communication error!
An abnormality occurred during instrument-internal communication.
206
Internal communication error!
An abnormality occurred during instrument-internal communication.
207
Internal Temperature Over!
Internal temperature is abnormally high. Automatically shuts down
after 10 seconds from occurrence.
208
Monochromator fan motor stopped!
The monochromator fan stopped.
209
PD temperature over!
Enterd emulation mode because there was a problem with the light
detector temperature.
210
Internal Temperature warning!
Warning for rising internal temperature.
211
Auto offset error!
Obtained an abnormal value during AUTO OFFSET operation.
212
Auto offset error!
Obtained an abnormal value during AUTO OFFSET operation.
213
Auto temperature control error!
Problem with the light detector temperature control.
214
Measurement sequence error!
The sweep stopped because measurement sequence fell into
disorder during the sweep.
220
Boot sequence error!
Started up in emulation mode because an abnormality occurred
upon start-up
221
Boot sequence error!
Started up in emulation mode because an abnormality occurred
upon start-up
222
Emulation Mode.
Entered emulation mode because an abnormality occurred.
223
Boot sequence error!
Started up in emulation mode because an abnormality occurred
upon start-up
224
Internal communication error!
An abnormality occurred during instrument-internal communication.
225
Internal communication error!
An abnormality occurred during instrument-internal communication.
226
Internal communication error!
An abnormality occurred during instrument-internal communication.
227
Internal communication error!
An abnormality occurred during instrument-internal communication.
228
Memory allocation error!
Data memory initialization failed.
230
Monochromator error!
Entered emulation mode because an abnormality with the operation
of the monochromator occurred.
231
Monochromator error!
Entered emulation mode because an abnormality with the operation
of the monochromator occurred.
232
Monochromator error!
Entered emulation mode because an abnormality with the operation
of the monochromator occurred.
233
Monochromator error!
Entered emulation mode because an abnormality with the operation
of the monochromator occurred.
234
Monochromator error!
Entered emulation mode because an abnormality with the operation
of the monochromator occurred.
235
Monochromator error!
Entered emulation mode because an abnormality with the operation
of the monochromator occurred.
236
Monochromator error!
Entered emulation mode because an abnormality with the operation
of the monochromator occurred.