5.4 Sweep Average
Procedure
Selecting a Trace to Average
1.
Press
TRACE
. The soft key menu for trace settings appears.
2.
Press the
ACTIVE TRACE
soft key. Traces A through G appear in soft keys.
3.
Press the soft key corresponding to the trace you wish to average. The selected
trace is set as the active trace, and becomes the target of the operation below.
4.
Press the
VIEW
soft key of the selected trace and select
DISP
. Each time you
press the soft key
DISP
switches to
BLANK
, or viceversa. (Trace B is given as an
example.)
Setting the Averaging Times
5.
Press the
ROLL AVE
soft key. The averaging times dialog box is displayed.
6.
Enter a number of averaging times using the rotary knob, step keys, or numeric
key pad.
7.
Perform measurement. The sweep average value is updated each time
measurement is performed.
note
The allowed setting for averaging times is 2-100.