5-52
IM AQ6376-01EN
Explanation
The instrument detects the peak (the maximum level value) or bottom (the minimum level
value) of the measured waveform.
PEAK SEARCH
Executes a peak search (a search for a maximum level value) on the active trace
waveform. A moving marker is placed, and the marker value is displayed in the data
area. If the peak level is above the screen top or below the screen bottom, a marker is
displayed at the screen top or bottom, but the actual (correct) marker value is displayed
in the data area. After measurement the marker can be moved with the rotary knob. Or, if
you press an arrow key, you can scroll the data area.
BOTTOM SEARCH
Executes a bottom search (a search for a minimum level value) on the active trace
waveform. A moving marker is placed, and the marker value is displayed in the data
area. If the bottom level is above the screen top or below the screen bottom, a marker is
displayed at the screen top or bottom, but the actual (correct) marker value is displayed
in the data area. After measurement the marker can be moved with the rotary knob. Or, if
you press an arrow key, you can scroll the data area.
NEXT LEVEL SEARCH
Sets a moving marker at the peak (maximum level value) or bottom (minimum level
value) which follows the currently set moving marker value (level value) in the active
trace waveform. If there is no such peak or bottom, a warning data is displayed.
WARNING 103 : No data in active trace
NEXT SEARCH RIGHT
Sets a moving marker at the peak (maximum level value) or bottom (minimum level
value) to the right of the currently set moving marker value (level value) in the active
trace waveform. If there is no such peak or bottom, a warning data is displayed.
WARNING 103 : No data in active trace
NEXT SEARCH LEFT
Sets a moving marker at the peak (maximum level value) or bottom (minimum level
value) to the left of the currently set moving marker value (level value) in the active trace
waveform. If there is no such peak or bottom, a warning data is displayed.
WARNING 103 : No data in active trace
SET MARKER
SET
Sets the moving marker as a fixed marker with the specified number.
A number from 001 to 1024 can be specified. The default is one greater than the highest
fixed marker number among the currently set markers, or the number 001 if no markers
have been set). If the MARKER ACTIVE soft key turns OFF, the SET MARKER soft key
is disabled.
CLEAR MARKER
CLEAR
Clears the specified fixed marker number. The marker value in the data area is also
cleared. The fixed marker number to be cleared (default value) is the last set fixed
marker number.
5.13 Single Search