8-24
IM AQ6376-01EN
Remote Monitor
This function can be used to monitor the instrument's screen and control the instrument
from a remote PC that is connected to the instrument through a TCP/IP port.
To use this function, you need remote monitoring software (not included). You cannot use
the remote monitoring port to perform remote control with normal remote commands.
User Name and Password
A user name and password are required to access the instrument using this function.
Use the menu that is displayed when you press the REMOTE USER ACCOUNT soft key
in the Ethernet setup to specify these settings. See the remote control user's manual, IM
AQ6376-17EN.
MONITOR PORT
Enables and disables the remote monitor TCP/IP port.
If you set this to OFF, the remote monitor is disabled.
PORT NO.
This is the remote monitor TCP/IP port number. It is fixed to 20001. You cannot use this
port to perform remote control with normal remote commands.
DISCONNECT
During remote monitoring, press this soft key to disconnect the instrument from the
remote PC.
This soft key can only be used when the instrument is connected to a PC.
Sharing Directories
The user area directory of the AQ6376 internal memory can be shared on a PC. When
the user area directory is shared, the files in the directory can be copied to the PC over
the network. You cannot save files to the AQ6376.
Turning the Display OFF
This function can be used to temporarily turn the display OFF. Use this function when
you are using the instrument in a darkroom or similar environment in which the light from
the screen has an effect on the work.
DISPLAY OFF
Press this soft key to turn OFF the backlight, which will also turn the display OFF.
When Remote Commands Have Been Used to Turn the Display OFF
If remote commands have been used to turn the display OFF, even if you perform panel
key or mouse operations to turn the display ON, the following message will be displayed
and the display will be turned OFF after approximately 5 seconds.
To keep the display turned ON, use a remote command, or press LOCAL to switch the
instrument from remote to local mode.
8.6 Other Settings