6.11 Measurement of Level Fluctuations in Single-
Wavelength Light (0 nm Sweeping)
Procedure
This function measures changes over time in the level of a specific wavelength level. It is
useful for purposes such as optical axis alignment when connecting an optical fiber to a
light source. The following discussion pertains to an example in which the spatial light of
an He-Ne gas laser (1523.488 nm) is input to an optical fiber.
Setting the Center Wavelength to 1523.488 nm
1.
Press
CENTER
. The soft key menu for setting the center wavelength appears.
2.
Press the
CENTER WL
soft key. The center wavelength setting screen is
displayed.
3.
Enter a center wavelength of 1523.488 nm using the rotary knob or numeric key
pad.
4.
Press
nm/ENTER
.
Setting the Resolution to 2.000 nm
5.
Press
SETUP
. The soft key menu for sweep condition settings appears.
6.
Press the
RESOLUTION
soft key. The resolution selection menu is displayed.
7.
Press the
2.000nm
soft key.
8.
Press
nm/ENTER
.
Setting the Sweep Width to 0 nm
9.
Press
SPAN
. The soft key menu for setting the sweep width appears.
10.
Press the
SPAN WL
soft key. The sweep width setting screen is displayed.
11.
Enter a sweep width of 0 nm using the rotary knob or numeric key pad.
12.
Press
nm/ENTER
. Set the sweep width to 0 nm. The measurement start
wavelength, measurement center wavelength, and measurement stop wavelength
are all set to 1523.488 nm.
Setting the Sweep Time
13.
Press the
0 nm SWEEP TIME
soft key. A screen for specifying the sweep time is
displayed.
14.
Enter a numerical value using the rotary knob or numeric key pad, then press
nm/ENTER
.
15.
Press
SWEEP
, followed by the
REPEAT
soft key. Sweeping begins.