4-38
IM AQ6376-01EN
4.15 Pulse Light Measurement
Procedure
The following three methods are available for measuring pulse light.
• Measurement using Peak hold mode
• Measurement as a time average spectrum (see “Explanation”).
• Measurement using External trigger mode.
Pulse Light Measurement Settings
1.
Press
SETUP
. The soft key menu for sweep condition settings appears.
2.
Press the
MORE
soft key. The MORE 2/2 soft key menu is displayed.
3.
Press the
PLS LIGHT MEASURE
soft key.
4.
Press the
PEAK HOLD
soft key. The peak hold value setting screen is displayed.
5.
Enter a peak hold value using the rotary knob, arrow keys, or numeric key pad. A
value larger than the period of the pulse light being measured must be entered for
the peak hold value.
6.
Press
ENTER
.
Displaying the Pulse Light Measurement Waveform
7.
Press
SWEEP
. The soft key menu for sweep appears.
8.
Press the
SINGLE
or
REPEAT
soft key. Sweeping begins, and a waveform is
displayed.
9.
To stop the sweep, press the
STOP
soft key.
Note
• An appropriate sensitivity is determined from the pulse width of the measured pulse light.
For details, see the table, “Sensitivity Name and Corresponding Pulse Width.”
• Enter a value longer than the period of the pulse light being measured for the peak hold
value.