5-32
IM AQ6376-01EN
Setting Measurement Sweep Width & Display Sweep Width with Line Markers
Setting the Measurement Sweep Width between Line Markers 1 and 2
<<See section 4.6 for details>>
With line marker 1 and 2 displayed, press the
MRK L1-L2 ->SPAN
soft key. The
sweep width setting screen and setting value are displayed. The measurement
sweep width, measurement start wavelength, and measurement stop wavelength
are changed. The measurement sweep width can be set by continuing in the DATA
ENTRY section. The available setting range is 0.5 to 1900 nm (in 1 nm steps).
Setting the Area from Line Marker 1 to Line Marker 2 as the Display Scale
ZOOM SPAN
With line marker 1 and 2 displayed, press the
MRK L1-L2 ->ZOOM SPAN
soft key.
The display sweep width setting screen and setting value are displayed. You can
also rewrite the currently displayed waveform according to the specified ZOOM
SPAN. See section 5.1.
Note
• If only one of the line markers is displayed, for L1, the wavelength on the right edge of the
screen is set to the measurement stop wavelength. For L2, the wavelength on the left edge
of the screen is set to the measurement start wavelength.
• The MKR L1-L2 ->ZOOM SPAN soft key cannot be used under the following conditions.
• When both L1 and L2 are OFF.
• When both SPLIT screens are on HOLD.
• When the span of the active trace is 0 nm.
Note
For information on analysis inside the line markers, see section 6.13, “Specifying an Analysis
Range.”
5.9 Marker Display