Villa Sistemi Medicali
VISITOR T15 – Installation & Acceptance -2
[File: M_30050B00.doc]
Rev.3 - Page 9/20
2.
2.
2.
2. Check the proper arm movement and
balancing in every operative condition, as
shown in the Fig. 6.
Verificare la corretta escursione e
bilanciatura del braccio in tutte le
condizioni operative, come illustrato in Fig.
Fig. 6
3.
3.
3.
3. Check the proper rotation and balancing
of the monobloc-collimator group in
respect of the transversal arm axis between
–18° and +133° (see Fig. 7).
Verificare la corretta rotazione e
bilanciatura del gruppo monoblocco-
collimatore rispetto l’asse trasversale del
braccio tra –18° e +133° (vedi Fig. 7).
Fig. 7