MNL000423.doc
Revision 01
09/18/2018
© Versum Materials, Inc. as of the revision and date shown. All rights reserved.
Confidential and Proprietary Data
Page
-
29
5.5.4
Level Check
–
Quartz Bubbler
For a 1.5L quartz bubbler, the level should read 100% if it is filled to 1200cc. A
template has been provided at the end of this section for testing the level sense probes
with the bubbler removed. A Template must be cut out and traced over with a
thicker/darker card stock material.
Inserting the card stock into the Source Cup Assembly and sliding across the receiver
window will simulate level changes in the bubbler. Each level must be present for at
least 5 seconds before the display will update to the new level because of the
discriminator function block. For the template, reference figure 5.9.