If the settings are not saved at this point the new calibration parameters will
be lost. The settings will default back to the original settings prior to
calibration.
1.2.4
Analog Output Calibration Procedure
The electronics chosen in the new design were specified for low drift due to time or
temperature. The factory calibration in most cases should not to be adjusted for long
periods of time. Should the output signals need calibration, the user must install the
Remote Display Module,
PN 152235
connected to the Power Supply Module at J3 to
complete. Contact your local Versum Materials, Inc. representative for more details on
Remote Display Module,
PN 152235
.
1.2.5
Voltage Output Calibration
1.
Remove the DDC connector from the Power Supply Module. Install the Remote
Display Module to J3 connector of the Power Supply Module.
2.
Connect the DMM positive to J5-7 and the common to J5-10 to check the
temperature output. Set DMM to DC voltage mode on 20 volt range.
3.
With the temperature probe uninstalled, the output should be 5.00VDC±0.05. If
the measurement is within specification no calibration is required.
4.
If the output requires calibration, navigate to the “CURR&VOLT CONFIG” menu
and press the green “√”.
5.
Press “√” at the “LEVEL & TEMP VOLT?” screen.
NOTE:
Entering this menu will cause the output to float high. The real reading will
return once you exit this screen.
6.
The “L” on the top row designates the calibration values for the voltage level
analog output. The value for percent can be adjusted to verify the output voltage
measurement. Likewise, the value next to the “T” can be adjusted to verify the