bubbler as required to achieve correct orientation. If the buzzer has been turned off,
this feature will not work. See ”Config Sys” to enable buzzer.
There are some quick steps from the main menu to assist the user. Press the red X
button to mute the alarm for the defaulted time duration. Press the right arrow once
and press the center green check button. Press the down arrow repeatedly until the
screen prompts to “CLEAR ALARM”. Press the center button again and the alarm will
be cleared from the display. If alarm condition is gone, press the Start/Stop switch to
the “START” position. The green LED will illuminate on the rocker switch.
A “High Temp” or “Low Temp” alarm may occur if the bubbler chemical temperature is
higher or lower than the default setting for set point tolerances. These alarms can be
cleared by changing the set point to a temperature different from the current setting or
changing the tolerance for the alarm condition.