Initializing pumps
42
Operating instructions
Argument <n2> options:
Argument <n3> options:
Distribution valve ports are numbered 1...X starting in a CW direction from
the syringe port. This is shown in Figures 5-14 (page 55) and 5-15 (page 56).
Initialize pump, set valve output to the left (CCW polarity)
(Y<n1,n2,n3>)
The [Y] command initializes the plunger drive and sets a non-distribution
valve output to the left (as viewed from the front of the pump).
If a distribution valve is configured, <n2> and <n3> select the input and out-
put ports. Valve movement occurs in a CCW (counterclockwise) motion. If a
non-distribution valve is configured, <n2> and <n3> are ignored.
The [Y] command resets the velocity settings ([V], [v] and [c]), and acceler-
ation [L] to power-up defaults. The increment mode [N1 or N2] is preserved
after a [Y] command.
Arguments:
Argument <n1> options:
<n2> options
Description
0
Set the input initialization port to 1 (default).
1...X
Set the input initialization port to a non-default port. X = number of
ports on the valve.
<n3> options
Description
0
Set the output initialization port to X (default).
1...X
Set the output initialization port to a non-default port. X = number
of ports on the valve.
<n1>
0...40
Plunger stall force and speed.
<n2>
0...X
Set the initialization input port on distribution valves, numbered
counter-clockwise from syringe port. X = number of ports.
<n3>
0...X
Set the initialization output port on distribution valves, numbered
counter-clockwise from syringe port. X = number of ports.
(n1)
0
Default argument
(n2)
1
Default argument
(n3)
X
Default argument, X = number of ports
<n1> options
Description
0
Initializes at full plunger stall force at speed 11 (default).
Recommended for syringes 1.0 mL
and larger.
1
Initializes at half plunger stall
force at speed 11.
Recommended for 250
µ
L
and 500
µ
L
syringes.
2
Initializes at one third plunger stall
force at speed 11.
Recommended for 50
µ
L
and 100
µ
L
syringes.
3
Initializes at full plunger stall
force and at speed 16.
4
Initializes at full plunger stall
force and at speed 18.
5...9
Same as <n> = 0.
10...40
Initializes at full plunger stall
force and at speed 10...40.