Using EEPROM
82
Operating instructions
Calibrate encoder levels and store in EEPROM (n<x>)
Note:
For factory use only.
Encoder voltage output levels can calibrated and set into EEPROM.
This command will run the pump through the calibration process, calculate
the values, and store them in EEPROM. These values can be read back using
the [?21] report command.
Table 5-11: Example full stroke variations based on [u] command selections
Configuration
Full stroke
normal mode (N0)
Full stroke
(N1 or N2 mode)
C3000/C3000MP
Set to step mode (default settings)
(u4_62, u12_0, u15_0)
3,000
24,000
C3000/C3000MP
Set to half-step mode
(u4_62,u12_1,u15_0)
6,000
48,000
C24000/C24000MP
1 mm lead screw pitch and half-step
mode (default settings)
(u4_248, u12_1, u15_1)
24,000
192,000