![TriContinent C24000 Product Manual Download Page 33](http://html.mh-extra.com/html/tricontinent/c24000/c24000_product-manual_1162143033.webp)
Install syringe
C-Series Product Manual
33
Install syringe
Note:
The C24000 pumps must be initialized before the syringe can be
installed. This is because these pumps cannot be moved manually. For
C24000 pumps, first go to "Initialize pump" (page 30) and return to this
procedure after initialization has been performed.
Refer to Figure 4-9 during this procedure.
1.
Lower the Syringe Plunger Arm/Holder to ensure sufficient room to
install the Syringe and Plunger.
–
C3000 pumps: if the pump is powered off, manually lower the
Plunger by pushing firmly down on the Syringe Plunger Arm/Holder.
If the power is on, this can be done by sending the command
[A3000R].
–
C24000 pumps: with the power on, send the command [A24000R].
The Plunger cannot be manually lowered on the C24000 pump due
to the high resolution of the drive system.
2.
Make sure the Plunger is fully seated in the Syringe, to ensure the length
is short enough to fit into the available space.
CAUTION:
–
Do not overtighten the Syringe or fittings. This can cause damage to
the Syringe and Valve, which can lead to leakage and reduced life.
Follow directions carefully.
–
Do not use Teflon® tape on Syringe or fitting threads. All Syringe
port seals are face seals (not thread sealing). Use of Teflon® tape can
lead to Valve damage due to over stress on the Valve seat from
excessive torque applied to the Syringe or fittings.
3.
Screw the threaded portion of the Syringe clockwise into the Valve (hold
the Syringe by the knurled portion when screwing it into the Valve).
4.
Tighten the Syringe until it bottoms in the Valve, then tighten another 1/4
turn (this roughly corresponds to 16 oz-in). If a more secure installation is
needed, loosen the Syringe and repeat the process approximately 24
hours later. The maximum torque specification is 21 oz-in.
5.
Pull the Plunger down and fully seat it into the Syringe Plunger Arm/
Holder.
6.
Securely tighten the Plunger Lock Screw by turning clockwise.
Summary of Contents for C24000
Page 8: ...This page intentionally blank ...
Page 14: ...This page intentionally blank ...
Page 142: ...This page intentionally blank ...