Plunger movements
62
Operating instructions
Make relative dispense (D<n>)
The [D] command moves the plunger upward the number of increments <n>
commanded. The new absolute position is the previous position minus <n>.
Arguments for C3000/C3000MP:
Arguments for C24000/C24000MP:
For example:
The syringe plunger is at position 3000. [D300] will move the plunger up
300 increments to an absolute position of 2700.
The [D] command will return an invalid operand error (error 3) if the final
plunger position would be less than 0.
Make relative dispense, not busy (d<n>)
This is the same as the [D] command, except that the status bit of the reply
string, and any subsequent query [Q] commands, indicates that the pump is
not busy.
<n>
0...3000
increments in N0 mode
<n>
0...24000
increments in N1 or N2 mode (micro-increments)
(0)
default argument
<n>
0...24000
increments in N0 mode
<n>
0...192000
increments in N1 or N2 mode (micro-increments)