Plunger movements
70
Operating instructions
Typically, this command is used during a dispense operation to provide an
early trigger to another pump to begin dispensing. This allows for continu-
ous dispense flow using two pumps.
For example, using the following string:
[J0IA3000j5007OA0GR]
Breakdown:
For a complete example of a continuous flow application, see "Dual pump
continuous flow application" (page 137).
Set syringe motor hold current (h<n>)
Note:
For factory use only.
Sets the syringe's motor hold, or non-moving, current to the value <n> spec-
ified in percent of maximum. On power-up, this value is overwritten to the
factory set default value.
Arguments:
Set syringe motor run current (m<n>)
Note:
For factory use only.
Sets the syringe motor run, or moving current, to the value <n> specified in
percent of maximum. On power-up, this value is overwritten to the factory
set default value.
Arguments:
J0
Set all 3 auxiliary outputs low.
I
Move valve to input position.
A3000 Aspirate: move syringe down to bottom of stroke.
j5007
Set all 3 auxiliary outputs high (binary equivalent of the number 7)
when syringe position is less than or equal to 500.
O
Move valve to output position.
A0
Dispense: move syringe up to top of stroke (position 0). Auxiliary
outputs will change state when syringe position is less than or
equal to 500.
G
Repeat continuously until terminated.
R
Execute command string.
<n>
0…100
% of maximum current
{10}
power-up default set at factory
<n>
0…100
% of maximum current
{75}
power-up default set at factory, C3000/C3000MP
{50}
power-up default set at factory, C24000/C24000MP