Dual pump continuous flow application
C-Series Product Manual
141
Command string breakdown for Pump B
s1
Store following string in EEPROM location 1 [s1].
J7ZV2000IA3000
Initial setup: set outputs high [J7], initialize pump
[Z], set velocity to 2000 increments/sec [V2000],
move valve to input position [I], aspirate a full
syringe [A3000] at set velocity.
Note:
for C24000 pumps, A3000 would be A24000
instead. This applies below also.)
OV1000
Prepare for dispense: move valve to outlet [O], set
dispense velocity to 1000 increments/sec
[V1000].
j100
When syringe position during dispense gets close
to top of stroke (10 in this case) [j10], set outputs
low [0]. This will trigger Pump A to begin its dis-
pense.
g
Start of continuous loop.
H0
Wait for trigger from Pump A.
A0
Dispense full stroke (trigger is sent to Pump A
when position is at 10, as stated above).
IV2000A3000J7O
V1000j100
When finished dispensing, move valve to input
position [I], refill [V2000A3000], and reset for
next dispense [J7OV1000j100].
G
Repeat forever (back to [g]).
R Run
command
string.