Command summaries
90
Operating instructions
Y<n1,n2,
n3>
0...40
0
Initialize plunger. Set valve to
the left on non-distribution
valves.
<n1> same as [Z<n1>]
<n2> same as [Z<n2>] except
counterclockwise
<n3> same as [Z<n3>] except
counterclockwise
W<n>
0...40
0
Initialize plunger without
valve.
Same as [Z<n1>]
z<n>
0...3000 N0
0...24000 N1/N2
(0...24000 N0
0...192000 N1/N2)
0
Set pump's internal position
counter to value specified.
k<n>
0…120 N0
0...960 N1/N2
24 (384)
24 (384)
Set syringe gap volume.
Plunger movement commands
A<n>
0...3000 N0
0...24000 N1/N2
(0...24000 N0
0...192000 N1/N2)
0
Move plunger to absolute
position.
a<n>
0
Move plunger to absolute
position, not busy.
P<n>
0
Relative pickup
p<n>
0
Relative pickup, not busy
D<n>
0
Relative dispense
d<n>
0
Relative dispense, not busy
Valve commands
I
N/A
N/A
Move valve to input position.
O
N/A
N/A
Move valve to output position.
B
N/A
N/A
Move valve to bypass position.
E
N/A
N/A
Move valve to extra position.
(Valid for 4-position valves
only.)
Set commands
K<n>
0…100
10 (80)
10 (80)
Set backlash increments.
L<n>
1...20 N0/N1
8...160 N2
14
Set acceleration/deceleration
slope.
Slope code
v<n>
1...1000 N0/N1
1...8000 N2
900
Set start velocity in
increments/sec.
V<n>
1...6000 N0/N1
1...48000 N2
1400
(5600)
Set top velocity in
increments/sec.
S<n>
0...40
11
11
Set a defined speed.
Speed code
c<n>
1...2700 N0/N1
1...21600 N2
900
Set cutoff velocity in
increments/sec.
C<n>
0...25
0
Set cutoff in increments.
Table 5-14: RS232/RS485 command summary (continued)
Command
Operand <n> range
( ) = C24000/
C24000MP only
Default
0perand
Power-up
default
Command description
Operand description