Connect tubing
Installation instructions
34
Figure 4-9. Installing syringe
Connect tubing
CAUTION:
Be sure to check fluid compatibility with the materials in the
valve, syringe, and plunger seal. Refer to "Chemical compatibility chart"
(page 130).
For fluid connections, the standard C-Series valves are equipped with
1/4-28" ports for the tubing fittings.
Procedure
Connect tubing to pump and fluid source and outlet according to your
instrument design and the tubing supplier’s instructions.
Note:
If more information is needed about valve port options, refer to “Valve
Positioning” and the associated illustrations in Chapter 5, Operating instruc-
tions (page 35).
The following application information may also be helpful:
•
"Optimizing pump performance" (page 133)
•
"Optimizing tubing performance" (page 135)
•
"Precision and accuracy" (page 131)
Valve
Syringe
Plunger
Screw Syringe clockwise to
seat in Valve body.
Pull Plunger down all the way
into Syringe Plunger Arm/Holder.
Turn Plunger Lock Screw
clockwise to tighten.
Plunger Lock Screw
Syringe Plunger Arm/Holder