Replacing components
97
C-Series Product Manual
CAUTION:
–
Do not overtighten the Syringe or fittings. This can cause damage to
the Syringe and Valve, which can lead to leakage and reduced life.
Follow directions carefully.
–
Do not use Teflon® tape on Syringe or fitting threads. All Syringe port
seals are face seals (not thread sealing). Use of Teflon® tape can lead
to valve damage due to over stress on the valve seat from excessive
torque applied to the Syringe or fittings.
3.
Screw the threaded portion of the Syringe clockwise into the Valve (hold
the Syringe by the knurled portion when screwing it into the Valve).
4.
Tighten the Syringe until it bottoms in the Valve, then tighten another 1/4
turn (this roughly corresponds to 16 oz-in). If a more secure installation is
needed, loosen the Syringe and repeat the process approximately 24
hours later. The maximum torque specification is 21 oz-in.
5.
Pull the Plunger down and fully seat it into the Syringe Plunger Arm/
Holder.
6.
Securely tighten the Plunger Lock Screw by turning clockwise.
Figure 6-17. Installing syringe
Valve
Syringe
Plunger
Screw Syringe clockwise to
seat in Valve body.
Pull Plunger down all the way
into Syringe Plunger Arm/Holder.
Turn Plunger Lock Screw
clockwise to tighten.
Plunger Lock Screw
Syringe Plunger Arm/Holder