Calculating flow rates
Reference information
128
Examples
•
1 mL syringe on C3000 in N0 or N1 mode at velocity [V] of 6000 incre-
ments/sec: flow rate = (1000
µL
/6000) * 6000 = 1000
µ
L/sec (see note
below)
•
1 mL syringe on C3000 in N2 mode at velocity [V] of 6000 increments/
sec: flow rate = (1000
µL
/48000) * 6000 = 125
µ
L/sec
•
1 mL syringe on C24000 in N0 or N1 mode at velocity [V] of 6000 incre-
ments/sec: flow rate = (1000
µ
L
/24000) * 6000 = 250
µ
L/sec
•
1 mL syringe on C24000 in N2 mode at velocity [V] of 6000 increments/
sec: flow rate = (1000
µ
L/192000) * 6000 = 31.25
µ
L
/sec
Note:
Velocity resolution is twice the position resolution for the C3000
product.
CAUTION:
Due to the smaller internal port diameters of the 6-way
distribution valve, it is recommended that fluid velocity does not exceed
1500
µ
L/sec when using this valve. Use caution when setting velocity with
2.5 mL and larger syringe sizes.Take these precautions to prevent cavitation
and excessive syringe back pressure.