Dual pump continuous flow application
Reference information
138
Tubing hookup
The fluid outputs of the two pumps are connected together with a “Y”
fitting. The inputs are independent.
Wiring
Special cable is required (TCS 8246-73). As shown below, this cable con-
nects an auxiliary output of each pump to an auxiliary input of the other
pump.
Jumper and address switch settings
Pump A is set to address 1, and Pump B is set to address 2 using the
address switches on the back of the two pumps (address 1 = switch posi-
tion 0, address 2 = switch position 1).
Pump A
Combined
continuous
flow output
Separate
inputs
Pump B
1
J3
IN
OUT
1
Power and communication cabling not shown
OUT
J3
IN
13
7
13
7
Pump B
Pump A