![TriContinent C24000 Product Manual Download Page 121](http://html.mh-extra.com/html/tricontinent/c24000/c24000_product-manual_1162143121.webp)
CAN interface communications
C-Series Product Manual
121
Common command example
After the host has sent command [A1000A0] to the pump, it can send
command 0 of frame type 2 to a pump and initiate execution of the buf-
fered command string. The pump here has been set to address 0.
1.
Host sends:
2.
Pump acknowledges:
3.
After executing the command, pump reports status:
Query command examples
Example: the host sends report command 29 of frame type 6 to a pump,
with the pump set to address 1.
1.
Host sends:
2.
Pump reports:
Note:
For query commands, no acknowledge frame is needed.
Direction
Group
Device
Frame
Type
RTR/IDE/r0
DLC
Data bytes
0
010
0000
010
000
0001
0x30 (“0”)
Direction
Group
Device
Frame
Type
RTR/IDE/r0
DLC
1
010
0000
010
000
0000
Direction
Group
Device
Frame
Type
RTR/IDE/r0
DLC
Data bytes
1
010
0000
010
000
0010
0x60 0x00h (“
`
0”)
Direction
Group
Device
Frame
Type
RTR/IDE/r0
DLC
Data bytes
0
010
0001
110
000
0010
0x32 0x39 (“29”)
Direction
Group
Device
Frame
Type
RTR/IDE/r0
DLC
Data bytes
1
010
0001
110
000
0010
0x60 0x00h (“
`
0”)