Set configuration jumpers and address switch
Installation instructions
22
Set configuration jumpers and address switch
Configuration jumpers
Before operating the pump, jumpers J2 and J9 must be set as appro-
priate for your system. Refer to the tables below and Figure 4-4.
Address switch setting
The address switch is normally used to assign the device address on the
RS485 bus (see Figure 4-4). There are 15 possible addresses (0-E hex).
The actual device address equals the address switch setting plus one. If
the address switch is set to 1, then, the device address is 2.
Position F is used for the self test. Use a small flat-bladed screwdriver to
rotate the switch. See cabling diagrams for examples.
Note:
The address switch function changes if AutoRun jumper J2-2 is
installed (see Table 4-2, above). For initial setup and confirmation, make sure
the AutoRun jumper is not installed.
Table 4-2: J2, various options
Jumper
Segment
Function
Action
1
Reserved
None. Spare jumper placed on this segment at factory.
2
AutoRun
Installed = AutoRun
1
Remove if unused. Recommend removing until operation has been con-
firmed. AutoRun executes string in EEPROM at power-up.
3
Overload detection
Installed = disabled
4
RS232/485 communica-
tion rate
Installed = 38,400 baud
Removed = 9600 baud
5
Valve type in use
Installed = 120
°
3-port valves
Removed = 90
°
4-port valves and all other valve types
2
1.
At power-up, if the AutoRun jumper is installed, the pump will execute the string at the memory location pointed
to by the address switch setting.
2.
All other valve types are EEPROM configurable. This is described in Table 5-6 (page 46).
Table 4-3: J9, RS485 and CAN terminations
1
Jumper
Segment
Function
Action
2
Application
1
RS485 A termination
Installed = terminated
Remove if unused.
Remove all termination
jumpers except on last
device on relevant bus. See
cabling diagrams for exam-
ples (beginning on page 26).
2
RS485 B termination
Installed = terminated
Remove if unused.
3
CAN termination
Installed = terminated
Remove if unused.
1.
In single-pump installations, the pump is always terminated.
2.
Remove, if desired, before mounting pump in a manner in which the termination jumper will be
obscured.