![TriContinent C24000 Product Manual Download Page 24](http://html.mh-extra.com/html/tricontinent/c24000/c24000_product-manual_1162143024.webp)
Make electrical connections
Installation instructions
24
Make electrical connections
Connector J3 pin assignments
All electrical connections, including power, communication, and control
signal IO are made through 15-pin D-sub connector J3. This connector is
located on the back of the pump. Mating connector and shield informa-
tion is provided in the tables below.
Cables are available for these connections. Contact Customer Service
for assistance.
Table 4-4: Mating connector and cover for J3
Item
Manufacturer
Description
Manufacturer's
part number
Mating
Connector
TE/TYCO
15-pin female solder cup,
receptacle
747909-2
Cover
TE/TYCO
Plastic cover (shield)
with male screw retainers
5207908-4
Table 4-5: Pin assignments for J3
Pin
Function
Details
1
24 VDC
24 VDC nominal, ±10%, 1.5A peak
2
RS232 TxD line
Output data from host
3
RS232 RxD line
Input data to host
4
Not used
5
CAN HI signal line
6
CAN LO signal line
7
AUX Input #1
TTL level (4.7K pullup to +5V)
8
AUX Input #2
TTL level (4.7K pullup to +5V)
9
Ground
Power and logic
10
Ground
Power and logic
11
RS485 A line
12
RS485 B line
13
AUX Output #1
TTL Level (1K pullup. Max source/sink current = ±20mA)
14
AUX Output #2
TTL Level (1K pullup. Max source/sink current = ±20mA)
15
AUX Output #3
TTL Level (1K pullup. Max source/sink current = ±20mA)
Mating
connector
and cover
J3
1
9
J3
CAUTION:
Make sure power is
off before making connections to
J3. Connecting or disconnecting
with power applied may result in
damage to equipment.