![TriContinent C24000 Product Manual Download Page 120](http://html.mh-extra.com/html/tricontinent/c24000/c24000_product-manual_1162143120.webp)
CAN interface communications
Reference information
120
Multi-frame command example
The host sends the command string [Z2S31A3000gHD300G10G5R] to a
pump, where the pump is set to address 0.
1.
Host sends in three consecutive messages:
Note:
The last frame type of a multi-message command is the type of com-
mand being sent. This example is an action command. Therefore the last
frame type is type 1. The first message in a multi-message command is type 3
and messages between the first and last are type 4.
2.
Pump acknowledges:
3.
After executing the command, the pump reports status:
Note:
“
`
0” decodes as first byte status = idle, followed by a null character
(hex zero). For multi-frame commands, the pump only sends an acknowledg-
ment for the last frame.
Direction
Group
Device
Frame
Type
RTR/IDE/r0
DLC
Data bytes
0
010
0000
011
000
1000
0x5A 0x32 0x53 0x33
0x31 0x41 0x33 0x30
(“Z2S31A30”)
Direction
Group
Device
Frame
Type
RTR/IDE/r0
DLC
Data bytes
0
010
0000
100
000
1000
0x30 0x30 0x67 0x48
0x44 0x33 0x30 0x30
(“00gHD300”)
Direction
Group
Device
Frame
Type
RTR/IDE/r0
DLC
Data bytes
0
010
0000
001
000
0110
0x47 0x31 0x30 0x47
0x35 0x52
(“G10G5R”)
Direction
Group
Device
Frame
Type
RTR/IDE/r0
DLC
1
010
0000
001
000
0000
Direction
Group
Device
Frame
Type
RTR/IDE/r0
DLC
Data bytes
1
010
0000
001
000
0010
0x60 0x00h (“
`
0”)