Set configuration jumpers and address switch
C-Series Product Manual
23
CAN baud rate
Default is 100K bps. For initial installation and operation, this should be
left unchanged. Once communication has been established, it may be
changed as described in Table 5-9 (page 79).
Figure 4-4. Configuration jumpers and address switch
Address Switch
REAR
VIEW
BOTTOM
VIEW
AutoRun
Overload Detection
Valve Type
Baud Rate
Not used
RS-485A
RS-485B
CAN
1
2
3
4
5
1
2
3
J2
J9
Do not plug or unplug DB-15
connector with power applied
or damage may result.
Installed = disabled
Installed = AutoRun
Installed = 38,400 baud
Removed = 9600 baud
Installed = 120° 3-port valve
Removed = All other valve types
Jumper installed =
termination active
Device address =
switch s 1.
Position F is for
self-test function.