Using EEPROM
79
C-Series Product Manual
Set pump configuration EEPROM parameters (U<n>)
These parameters are only read on power-up. Thus they will take effect only
when the power is cycled.
This command does not require an [R] to execute. Arguments for the [U]
command are shown in Table 5-9.
Table 5-9: Pump configuration set with U<n>
<n> value
Description
0
Reserved for future use.
1
3-port 120° Y valve
Valve selection jumper on the back of the pump must also be installed.
2
4-port 90° valve
Valve selection jumper on the back of the pump must also be removed.
3
Reserved for future use.
4
3-way distribution valve that uses [I], [O], [B] and [E] commands.
5
3-port and 4-port T valves
7
6-way distribution valve
8
Reserved for future use.
9
4-port loop valve
11
3-way distribution valves that use [I]<n> and [O]<n> commands
30
Enable AutoRun mode.
This setting performs the same function as the AutoRun jumper on the
back of the pump.
Note:
If the AutoRun jumper is installed, it will override the EEPROM
setting.
31
Clear AutoRun mode.
Note:
If the AutoRun jumper is installed, it will override the EEPROM
setting.
41
Reserved for future use.
47
Reserved for future use.
51
Set CAN baud rate to 100K bps.
52
Set CAN baud rate to 250K bps.
53
Set CAN baud rate to 500K bps.
54
Set CAN baud rate to 1M bps.
57
Set CAN baud rate to 125K.