![TriContinent C24000 Product Manual Download Page 55](http://html.mh-extra.com/html/tricontinent/c24000/c24000_product-manual_1162143055.webp)
Valve movements
55
C-Series Product Manual
Figure 5-14. Positions for 3-way distribution valve (U11 config)
Syringe port located at bottom of valve
3-way distribution commands
using Z initialization
3-way distribution commands
using Y initialization
I <1> or O <1> connects
syringe to port 1
I <1> or O <1> connects
syringe to port 1
3
2
1
O <2> or I <2> connects
syringe to port 2
O <2> or I <2> connects
syringe to port 2
I <3> or O <3> connects
syringe to port 3
O <3> or I <3> connects
syringe to port 3
I= CW rotation
O= CCW rotation
Jumper removed for 90° valve
3
2
1
3
2
1
1
2
3
1
2
3
1
2
3