6. Troubleshooting
Cause 2
The connection setup of the instrument is not cor-
rect.
Remedy
Give correct connection setup in the setup
procedures mentioned in "3.11.2 Setting
items for B-scan diagnosis asistant function/
A-scan diaguosis function f) Setting of com-
munication". As for confirming on the details
of this setting, see the Operation Manual for
TOMEY Link and also ask for your System
Administrator.
z
z
z
z
z
"Invalid Patient ID!! (044)" is displayed.
Cause 1
Patient information corresponding to inputted ID
number is not registered to the TOMEY Link.
Remedy
Appropriate countermeasures shall be taken
by referring to the Operation Manual for
TOMEY Link, including registering of patient
information.
z
z
z
z
z
"Server Connection ERROR!! (042)" is displayed.
Cause 1
The setup of the instrument is not correct.
Remedy
Give correct connection setup in the setup
procedures mentioned in "3.11.2 Setting
items for B-scan diagnosis asistant function/
A-scan diaguosis function f) Setting of com-
munication". As for confirming on the details
of this setting, see the Operation Manual of
TOMEYLink and also consult your System
Administrator.
z
z
z
z
z
"Database ERROR!! (045)" is displayed.
Cause 1
ODBC setting on TOMEYLink is not correct.
Remedy
Setup the ODBC of TOMEYLink correctly.
Check the Operation Manual of TOMEYLink
and consult your System Administrator.
Summary of Contents for UD-600
Page 2: ......
Page 14: ...xii This page is intentionally blank...
Page 22: ...1 8 This page is intentionally blank...
Page 28: ...2 6 This page is intentionally blank...
Page 157: ...3 129 Simple mode 3 9 Printout...
Page 159: ...3 131 Simple mode 3 9 Printout...
Page 192: ...3 164 This page is intentionally blank...
Page 256: ...8 14 This page is intentionally blank...
Page 260: ...A 4 INDEX...
Page 262: ...0123 ID 0608...