8-13
8. Appendix
150 kHz to 80 MHz
d =1.2
P
Recommended separation distances between
portable and mobile RF communications equipment and the
UD-6000
Table 206
Rated maximum output
power of transmitter
W
The
UD-6000
is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the
UD-6000
can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications equipment (transmit-
ters) and the
UD-6000
as recommended below, according to the maximum output power of the communi-
cations equipment.
Separation distance according to frequency of transmitter
m
80 MHz to 800 MHz
d =1.2
800 MHz to 2.5 GHz
d =2.3
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P
is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufac-
turer.
Note 1: At 80MHz and 800MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Summary of Contents for UD-600
Page 2: ......
Page 14: ...xii This page is intentionally blank...
Page 22: ...1 8 This page is intentionally blank...
Page 28: ...2 6 This page is intentionally blank...
Page 157: ...3 129 Simple mode 3 9 Printout...
Page 159: ...3 131 Simple mode 3 9 Printout...
Page 192: ...3 164 This page is intentionally blank...
Page 256: ...8 14 This page is intentionally blank...
Page 260: ...A 4 INDEX...
Page 262: ...0123 ID 0608...